美国电影禁忌乱偷DVD版高清,特级毛片a级毛片免费播放,亚洲一区二区三区影院,亚洲AV无码专区在线观看下载

?

常州市衡正電子儀器有限公司是一家集軟件開發(fā)、制造、銷售、服務(wù)于一體的科技型現(xiàn)代化企業(yè)。
掃描手機二維碼
掃描微信二維碼
13861223798

走進衡正電子

INTO HENGZHENG ELECTRONICS

您當前所在位置:首頁 > Technical application網(wǎng)站公告:測試信息來自互聯(lián)網(wǎng),若涉及侵權(quán),請聯(lián)系我們刪除!

關(guān)于我們列表

企業(yè)文化
營銷網(wǎng)絡(luò)

新聞推薦

電子天平使用應(yīng)具備的基本知識

電子密度天平計量檢定項目

在日常的周期檢定和常規(guī)的產(chǎn)品質(zhì)量檢查中,一般需要檢定以下幾項內(nèi)容...

JA系列電子天平選購應(yīng)該注意哪...

JA系列電子天平對操作人員的兼容性和對操作環(huán)境的耐受能力。

如何正確使用FA系列電子天平?

準備將FA系列電子天平放在穩(wěn)定的工作臺上,避免振動,氣流、陽光直...

FA系列電子天平使用的注意事項

電子天平如何測量固體密度

鹵素水分測定儀操作過程

聯(lián)系我們

地址:江蘇省常州市鐘樓區(qū)龍城大道2188號新閘科技園

聯(lián)系人:查小姐
郵編:213000
電話:0519-83263587
傳真:0519-83268313
網(wǎng)址:http://n4636.cn/

1 open question _ no display after opening
(1) There is no 220V current on the power outlet to turn on the AC outlet
(2) AC adapter error
Choose the 220V ~ AC adapter (external transformer) suitable for our work
AB-S / A8-1.4v ~ 50 / 60HZ6VA
PB-S / A9.5-20VDC6W
AB / PB / GB9.5-14.5V50 / 60HZ1.5VA
(3) AC adapter burned
Replace the AC adapter
2 open the problem _ completely no display after opening
(1) Calibration data is missing
Recalculate the electronic balance
(2) instantaneous interference effects
3 open the problem _ open completely no display
Using the balance of the school, AB-S / A, PB-S / A
Using the balance of the outside school, AB / PB / GB / SB
(1) Electronic balance placed environment bad Windproof window is not closed
Improve the electronic balance of the placement environment, close all windproof windows
(2) calibration characteristics
AB-S / A balance 1 hour after the first warm-up, do the first automatic calibration; the second environment is open after the balance 2 hours; and then maintain the power balance after 150 hours, on-time automatic calibration if necessary, you can manually trigger Calibration (internal school).
The PB-S / A balance is automatically calibrated when it is turned on and, if necessary, the calibration can be triggered manually at any time. AB / PB / GB / SB can only be outside the school.
(3) calibration error
AB / PB / GB / SB with the wrong, the external weight of the school. When entering the outside school program, the balance will appear a flashing weight value, the use of this value of the weight of the outside school.
(4) AB-S / APB-S / A balance display at the top right of the "CALL"
When the display appears "CALL", the monitor makes the balance is now working, and need to do the school. After completing the school "CALL" will automatically disappear.
4 open the problem _ completely no display after opening
(1) The electronic balance placed in the environment is too bad
Improve the environment
(2) The parameters in the electronic balance menu are not set well
After recalling the menu, use the "RESET" function to exit the menu correctly and return to the factory settings
(3) weighing the room to keep the body temperature
Minimize this human factor
(4) the temperature of the object to be weighed is not equal to the balance
Place the sample beside the balance.
(5) samples exist water absorption, water discharge, static electricity, magnetic ... ... and other characteristics.
Refer to the METTLER TOLEDO balance for proper weighing / electronic analytical balance to overcome these effects in order to properly weigh and properly use the microbalance.
5 open the problem _ open completely no display
(1) the balance placed in the environment is too bad
Environmental factors include: vibration, airflow, temperature, external magnetic field, must improve the environment, close the weighing room of the windshield, balance to work properly.
(2) The parameters in the balance menu are not set well
After the menu is transferred, use the "RESET" function to exit the menu correctly and return to the factory settings
6 open the problem _ completely no display after opening
1, the desiccant of the water and water to form a different direction of the air flow, causing the air buoyancy changes, resulting in weighing instability
The desiccant in the weighing chamber should be removed to maintain a stable weighing environment
7 open the problem _ completely no display after opening
(1) not loaded plate
After the power, first installed the correct weighing plate, and then open the balance
(2) said the disk wrong
Use the correct title plate to match the balance
(3) said the disk and the wind collision
Due to improper installation of the reasons, please find out the reasons for re-correct installation.
聯(lián)系我們
地址:常州市鐘樓區(qū)龍城大道2188號新閘科技園10棟313
聯(lián)系人:查小姐
電話:0519-83263587
傳真:0519-83268313
網(wǎng)址:http://www.czhz17.com
關(guān)注衡正
Copyright ? 2017 - 2024 All Rights Reserved 常州市衡正電子儀器有限公司 版權(quán)所有  蘇ICP備17024635號 技術(shù)支持:冉冉科技 網(wǎng)站地圖